Description
Capnograph CA10M true colour 1.3″ screen
Capnograph CA10M true colour 1.3″ screen
Description:
A Capnograph adopts mainstream methods to measure the end-tidal carbon dioxide, airway respiration rate and other parameters. It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow condition, as well as indirectly reflect the alveolar ventilation, which is applicable for use in pre-hospital emergency transfer, emergency department, ICU, operating room, neonatology, respiratory department and other occasions. The application population is adult, paediatric and neonate patients. The Capnograph connects with patient’ respiratory tract by the airway adapter and breathing tube.
Performance:
- End-tidal CO2 measurement:
- Measurement range: 0 ~150mmHg
- Resolution: 1mmHg
Accuracy:
- 0~40 mmHg:±2 mmHg
- 41~70 mmHg:±5% of readings
- 71~100 mmHg:±8% of readings
- 101~150 mmHg:±10% of readings
Respiration rate measurement:
- Measurement range: 2rpm~150rbpm
- Resolution: 1rpm
- Accuracy: ±1rpm
Safety characteristic:
- Device type: internally powered equipment
- Degree of protection against electric shock: type BF applied part
- Degree of protection against ingress of liquid: IPX2
Accessories:
- User Manual
- Adult gas path adapter
Specifications:
- Dimension: 55×52×59mm
- Weight: 97g (with batteries)